Year Established:20132013
Business Type:
Management Certifiation:
System Function
Type: Full automatic , random access
Though put: 400tests/hour, 800test with ISE
Assay tapes: End point, Kinetics, fixed time, immunoturbidimetry, Monochromatic, Bi-chromatic & Multi-standard Analysis Linearity & Nonlinearity Clibra Single & Dual Reagents can be used
Test Range: Clinical chemistry , Specific problem, ISE, Drug abuse, Urine, Cerbrospinal fluid, etc.
Sample System
Sample tray: 115 sample position
Sample Volume: 1.5 ~35μl (in steps of 0.1μl)
Sample probe: clot detection, liquid level detection
Dilution function
Reagent System
Reagent Tray: 90 reagent positions.
Reagent volume: 15μl~300μl
Barcode reader is optional
Reaction System
Reaction tray: 120 cuvette light pass length: 6mm
Reaction liquid volume: 150μl~300μl
Reaction temperature: 37℃±0.1℃
Optical System
Light source: Halogen lamp 6V/10W, light span≥3000 hours
Grating Measuring wavelength: 340 to 800nm, 12 wavelengths invariable
Detector: photodiode array
Rear spectrophotometry of holographic concave flat field grating
Calibration and Quality control
Assay tapes: End point, Kinetics , fixed time. immunoturbidimetry, Monochromatic,Bi-chromatic &
Multi-standard Analysis, Linearity & Nonlinearity Calibration Single & Dual reagents can be used
Calibration Mode: Linear (one-point, two point and mult-point), logit-log4p, logit-log5p nSpine, Exponential , Polynomial ,Parabola.
Calibration Period: setting as required.
OC Rule: Westgard Multi-rule, Cam μ Lative sum check and Twin plot
OC Mode: Intra day & Inter day
Working Condition:
System interface: Network port
Power Supply : 220V AC.50/60Hz
Power consumption: ≤1000VA
Environmental temperature: 15~30℃
Humidity: ≤85%
Dimension:1146X760X1095mm
Weight: 500kg